Recall Z-0557-2019 ยท OWB Interventional Fluoroscopic X-Ray System

Stephanix

STEPHANIX D2RS Digital Dynamic Remote Systemhttp://cts.fda.gov/division-tracking/images/trash.png Product The D2RS Digital Dynamic Remote System is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and intervent

DateNov 2, 2018
StatusTerminated
ReasonThe firm has detected a potential risk using the command. After releasing the command, the movement of the table may continue instead of stopping.
Root causeSoftware design
ActionThe firm initiated the recall by email on 11/02/2018 to the sole US distributor. This email included a letter to be provided to the end user. The letter directed the user as follows: "As a precautionary measure and before your usual maintenance team intervention, the user should look at the equipment during all of its movement (as stated in the user manual). Immediately release the command if an abnormal movement is detected. If necessary, use the closest emergency switch to stop the system." f you need more information or technical assistance, you can phone your usual contact.
Quantity17 units
DistributionUS Nationwide Distribution in the states of FL, IN, WV, TN, MO, WI, CA, FL, GA, PA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102529D2RF DIGITAL DYNAMIC REMOTE SYSTEM
Stephanix Radiological Systems, Oct 7, 2010