Recall Z-0486-2019 ยท IQA System, Environmental Control, Powered

Ascom (US) Inc.

Ascom Telligence Nurse Call System

DateNov 2, 2018
StatusOpen, Classified
ReasonThere is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.
Root causeSoftware Manufacturing/Software Deployment
ActionAscom Partners / Distributors were contacted by e-Mail, USPS Mail and Phone as necessary on 11/02/2018 to determine where the Telligence 5.0 Product is installed.
Quantity57 units
DistributionAL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall