Recall Z-0614-2019 ยท DQK Computer, Diagnostic, Programmable

Philips Electronics North America Corporation

Xper Flex Cardio Physiomonitoring System, Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 453564241911, 453564483321, 453564483331, 453564621791, 453564621801, 453564669081

DateNov 15, 2018
StatusTerminated
ReasonThere are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.
Root causeSoftware Design Change
ActionThe firm initiated the recall on X by letter. The letter explained the issues, potential risk, and provided workarounds for the consignee to use until the new software is ready for release.
Quantity651 units
Distributionworldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101571XPER INFORMATION MANAGEMENT AND FLEX CARDIO PHYSIOMONITORING SYSTEMS
Witt Biomedical Corp, Oct 26, 2010