Recall Z-0457-2019 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

ARTIS zee biplane systems with Material 10094141- a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee includes cardiac angiography, neuro angiography, general angiograph

DateOct 19, 2018
StatusTerminated
ReasonVibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
Root causeDevice Design
ActionSiemens issued on 10/19/18 an Urgent Medial Devoce Correction notification to via AX033/18/S advising users of the problem, health risk, steps to be taken. Siemens has initiated screw connection exchange of the ceiling bearing via Update Instruction AXOSS/17 /S.
Quantity14 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073290ARTIS ZEE AND ZEEGO
Siemens Medical Solutions USA, Inc., Feb 11, 2008