ClearanceBrief
Recall Z-0553-2019 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Zimmer Biomet, Inc.

Biomet 20mm Compress Device Short Anchor Plug With Drill Set / With Pin Inserter Ion Implant Item number: 178562 Product Usage: 1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. 2. Tumor resections. 3. Revision of previously failed total joint arthroplasty. 4. Trauma. The Compress Segmental Femoral Replacement System (Short Spindle and Anchor Plug) c

DateOct 24, 2018
StatusTerminated
ReasonPotential metal burrs in the holes of the device preventing the appropriate drill and transverse pins from being inserted
Root causeEmployee error
ActionZimmer Biomet initiated recall by letter on 10/24/18 to distributors and hospitals advising of the problem and action to be taken-locate quarantine and return product. Complete Inventory Return Certification Form. Questions or concerns call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.
Quantity6 units
DistributionUS Nationwide Distribution in the states of IN, DC, MD
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062998COMPRESS SEGMENTAL FEMORAL REPLACEMENT SYSTEM (SHORT SPINDLE AND ANCHOR PLUG)
Biomet Manufacturing Corp, Jan 19, 2007