Recall Z-0595-2019 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Angiodynamics Inc. (Navilyst Medical Inc.)

Xcela Power Injectable PICC with PASV (XCELA PASV 4F SL 55CM Catheter Kit CE PG), Cat no. 55-711

DateOct 16, 2018
StatusTerminated
ReasonSpecific lots of Xcela Power Injectable and BioFlo PICCs are labelled with incorrect expiration dates. The labelled expiration date extends the shelf life of the product beyond the date supported by validation testing. Although the product under recall has not yet reached its validated expiry date, removal of the affected product is warranted to prevent use beyond the validated expiry date.
Root causeUnder Investigation by firm
ActionOn October 16, 2018, the firm issued Urgent Voluntary Medical Device Recall letters to affected customers. The letter identified the product issue. Customers were asked to do take the following immediate actions: 1. Actions to be taken IMMEDIATELY: - Stop using the product subject to recall. - Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). - Segregate this product in a secure location for return to AngioDynamics, Inc. - Forward a copy of this recall notification to all sites to which you have dist
QuantityBox of 1 Units
DistributionDistributed to accounts in Florida. Foreign distribution to Great Britain.
FDA recordOpen on fda.gov