Recall Z-2292-2019 ยท OOI Real Time Nucleic Acid Amplification System

Luminex Corporation

Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories. Scanner Label/Labeling: JADAK JDK-2330 REV B 170512-017

DateAug 19, 2019 (initiated Jun 29, 2018)
StatusTerminated
ReasonAn incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order.
Root causeProcess control
ActionOn July 10, 2018, the firm sent an "URGENT: FIELD SAFETY NOTICE" to customers via email (with a follow-up telephone call) informing them that they may have received an affected barcode scanner. The Notice explained that when the potentially affected barcode scanner is used on multiple ARIES cassettes at the same time, the system may prepopulate the Sample ID field with a previously stored sample ID. This field will be overwritten by scanning or manually entering a sample, but in the event the ARIES cassette is not scanned, the prepopulated field may cause the incorrect association of sample id
Quantity153 (140 U.S. and 13 O.U.S.)
DistributionWorldwide distribution. US nationwide, Canada, Ireland, Netherlands, Singapore, Thailand, and United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151917ARIES System
Luminex Corporation, Oct 6, 2015
K160517ARIES System
Luminex Corporation, Apr 12, 2016
K161495ARIES MI System, ARIES HSV 1&2 Assay Cassettes-Carton of 24 (IVD), ARIES HSV 1&2 Assay Protocol File Kit (IVD)
Luminex Corporation, Jun 30, 2016
K162772ARIES GBS Assay
Luminex Corporation, Dec 23, 2016