Recall Z-0437-2019 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Maquet Cardiovascular Us Sales, Llc

QUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

DateNov 28, 2018 (initiated Oct 25, 2018)
StatusTerminated
ReasonThere is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
Root causeNonconforming Material/Component
ActionThe firm, GETTINGE, sent an "URGENT MEDICAL DEVICE RECALL" notification letters dated 10/25/18 to customers. The letters described the product, problem, and actions to be taken. The customers were instructed to do the following: -Please immediately examine your inventory to determine if you have any of the QUAD ROX-i Neonatal Oxygenator with the product codes/lot numbers listed in this notice. -If you wish to use the product, please remove and do not use the enclosed reducing screw adapter 1/4" x 3/16" (Ref. No. 70104.8593). -If you do not wish to use the product as packaged, please contact Ma
Quantity490 total in US
DistributionWorldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakist
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102464QUADROX-I NEONATAL MICROPOROUS MEMBRANE OXYGENATOR WITH INTEGRATED ARTERIAL FILTER WITH SOFTLINE/BIOLINE COATING
Maquet Cardiopulmonary, AG, Mar 4, 2011
K141432NEONATAL VENOUS HARDSHELL CARDIOTOMY RESERVIOR, PEDIATRIC VENOUS HARDSHELL CARDIOTOMY RESERVIOR, NEONATAL VENOUS HARDHSH
Maquet Cardiopulmonary, AG, Jul 7, 2014