Recall Z-0436-2019 ยท DTZ Oxygenator, Cardiopulmonary Bypass
Maquet Cardiovascular Us Sales, Llc
QUADROX-i Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055120 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.
| Date | Nov 28, 2018 (initiated Oct 25, 2018) |
| Status | Terminated |
| Reason | There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection. |
| Root cause | Nonconforming Material/Component |
| Action | The firm, GETTINGE, sent an "URGENT MEDICAL DEVICE RECALL" notification letters dated 10/25/18 to customers. The letters described the product, problem, and actions to be taken. The customers were instructed to do the following: -Please immediately examine your inventory to determine if you have any of the QUAD ROX-i Neonatal Oxygenator with the product codes/lot numbers listed in this notice. -If you wish to use the product, please remove and do not use the enclosed reducing screw adapter 1/4" x 3/16" (Ref. No. 70104.8593). -If you do not wish to use the product as packaged, please contact Ma |
| Quantity | 490 total in US |
| Distribution | Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakist |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K102464 | QUADROX-I NEONATAL MICROPOROUS MEMBRANE OXYGENATOR WITH INTEGRATED ARTERIAL FILTER WITH SOFTLINE/BIOLINE COATING Maquet Cardiopulmonary, AG, Mar 4, 2011 |
| K141432 | NEONATAL VENOUS HARDSHELL CARDIOTOMY RESERVIOR, PEDIATRIC VENOUS HARDSHELL CARDIOTOMY RESERVIOR, NEONATAL VENOUS HARDHSH Maquet Cardiopulmonary, AG, Jul 7, 2014 |