ClearanceBrief
Recall Z-0782-2019 ยท JIT Calibrator, Secondary

Siemens Healthcare Diagnostics, Inc.

Atellica CH 930 Analyzer. In vitro diagnostic testing of clinical specimens Siemens Material Number (SMN): 1 067000 - Product Usage: The Atellica Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated

DateSep 26, 2018
StatusTerminated
ReasonMultiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler Prime
Root causeSoftware design
ActionSiemens Healthcare Diagnsotics issue an Urgent Medical Device Correction (UMDC) was issued to US customers via Fed-Ex on September 26, 2018. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities. The letter explains the reasons, observed behaviors, risk to health, actions to be taken. Review the letter with the Medical Director. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. The system will be updat
Quantity490 units
DistributionWorldwide distributions - US Nationwide and countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, India, Israel, Italy, Japan, Latvia, Malaysia, Mexico, Netherlands, Nigeria, Norway, P.R. China,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151767Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Br
Siemens Healthcare Diagnostics, Inc., Dec 11, 2015