Recall Z-0561-2019 ยท JPB Hemoglobin M

Helena Laboratories, Inc.

Cascade Abrazo c-ACT-LR Test Card IVD FOR IN VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES 003366 12/13(2) FOR EXPORT ONLY

DateFeb 24, 2017
StatusTerminated
ReasonDue to some cards in this lot reported to have experienced variations in results, (verified by wave form), resulting in some cards with shorter or longer times than expected.
Root causeDevice Design
ActionOn 02/24/2017, the firm sent an email to affected users, with a Withdrawal Notification. The firm informed the customers/distributors that Due to some cards in this lot experiencing variations in wave form, resulting in some cards with shorter or longer times than expected. The firm instructed the customers/distributors to disposes of affected lot, as a preventive quality measure. On 04/12/2017, the firm mailed formal letters to all customers. PUBLIC CONTACT: "Name: Dr. Jessica Jones Hanka "Title: Technical Services Manager "Address: 1530 Lindbergh Drive, Beaumont, TX 77707 "Telephone #:
Quantity27 kits
DistributionUS: Products under control of Manufacturer in TX OUS: Four International Distributors - Australia, Netherlands, Italy, and England
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.