Recall Z-0951-2019 ยท MEB Pump, Infusion, Elastomeric
Arrow International Inc
AF 550x8ML/H 2ML 15LO 8.7" DUAL CATHS, 4, Product code MT8015XLYK10 CP, Arrow AutoFuser Disposable Pain Control Pump
| Date | Oct 2, 2018 |
| Status | Terminated |
| Reason | Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration. |
| Root cause | Process control |
| Action | On October 2, 2018, the firm notified consignees via an Urgent Medical Device Recall letter. The letter informed customers of the product issue. Customers were instructed to take the following actions: 1. If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Recall Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) number and pr |
| Quantity | 8 |
| Distribution | Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K090300 | MODIFICATION TO AUTOFUSER Ace Medical Us, LLC, Apr 30, 2009 |