ClearanceBrief
Recall Z-0942-2019 ยท MEB Pump, Infusion, Elastomeric

Arrow International Inc

AF550x5ML/H 2ML 60LO 6" CATH, 4.5" IN, Product code MT5060XLSK5 CP, Arrow AutoFuser Disposable Pain Control Pump

DateOct 2, 2018
StatusTerminated
ReasonPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Root causeProcess control
ActionOn October 2, 2018, the firm notified consignees via an Urgent Medical Device Recall letter. The letter informed customers of the product issue. Customers were instructed to take the following actions: 1. If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Recall Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) number and pr
Quantity8
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090300MODIFICATION TO AUTOFUSER
Ace Medical Us, LLC, Apr 30, 2009