Recall Z-0562-2019 ยท ITX Transducer, Ultrasonic, Diagnostic

CIVCO Medical Instruments Co., Inc.

CIVCO Needle Guide Bracket, REF numbers M001457100, P06013-10, and P03122-08, reusable non-sterile single-angle bracket for use with SonoSite L25 series tranducers, RX. The firm name on the label is CIVCO, Kalona, IA.

DateNov 2, 2018
StatusTerminated
ReasonThe needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.
Root causeOther
ActionThe recalling firm began issuing letters dated 10/30/2918 via email on 11/2/2018 requesting return of the boxes of product or reusable brackets that may be in their inventory.
Quantity738 boxes
DistributionForeign distribution was made to Canada, Argentina, Australia, Austria, Belgium, Brazil, Chile, China, Colombia, Croatia, Cyprus, Denmark, Estonia, France, Germany, Hong Kong, India, Israel, Italy, Japan, Latvia, Lebanon, Lithuania, Netherlands, New Zealand, Philippines, Poland, Portugal, Qatar, Rom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030064ILOOK 25 NEEDLE GUIDE ATTACHMENT AND BRACKET ASSEMBLY
Sonosite,Inc., Jan 22, 2003