ClearanceBrief
Recall Z-0413-2019 ยท BSZ Gas-Machine, Anesthesia

GE Healthcare, LLC

Touchscreen display kit. Sold under the following product names: TOUCHSCREEN DISPLAY KIT 11.X ENGLISH FOR USA, TOUCHSCREEN DISPLAY KIT 11.X ENGLISH, TOUCHSCREEN DISPLAY KIT 11.X GERMAN, TOUCHSCREEN DISPLAY KIT 11.X DUTCH, TOUCHSCREEN DISPLAY KIT 11.X FRENCH, TOUCHSCREEN DISPLAY KIT 11.X DANISH, TOUCHSCREEN DISPLAY KIT 11.X ESTONIAN, TOUCHSCREEN DISPLAY KIT 11.X ITALIAN, TOUCHSCREEN DISPLAY KIT

DateNov 9, 2018 (initiated Sep 27, 2018)
StatusTerminated
ReasonUpgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
Root causeSoftware design
ActionGE Healthcare notified customers on about 09/27/2018 via "URGENT MEDICAL DEVICE CORRECTION" letter. The letter informed customers that they could continue to use the anesthesia device. The issue can only occur if the patient stops spontaneous breathing in PSVPro ventilation mode after completion of lung recruitment breaths. Stopping the Cycling Procedure resolves the issue. Continue to monitor patient as is normal for Apnea. A GE Healthcare representative will contact each customer to arrange for the correction. Questions or concerns can be directed to GE Healthcare Service at 1-800-437
Quantity4,207 units total
DistributionWorldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Cze
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110213GE DATEX-OHMEDA AISYS
Datex-Ohmeda, Inc., Jul 20, 2011
K170872Aisys CS2
Datex-Ohmeda, Inc., Aug 15, 2017