Recall Z-0236-2019 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Medical Systems, LLC

SIGNA Architect system Product Usage: The Signa Architect system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/

DateOct 3, 2018
StatusTerminated
ReasonA small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
Root causeDevice Design
ActionOn about 10/03/2018, GE Healthcare sent URGENT MEDICAL DEVICE CORRECTION letters to their consignees informing them of recall and providing safety instructions. Consignees were instructed that they can continue to use the system by following the existing Operator Manual Safety Information for proper patient padding and observation during scanning. The safety information contained in the Operator's Manual includes: 1. WARNING: Place appropriate non-conductive padding between the patient and the bore wherever a portion of the body may come into contact with the magnet opening. 2. CAUTI
Quantity4 units
DistributionWorldwide - US Nationwide Distribution in the state of Utah. International distribution to France, Republic of Korea, and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160618Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, Optima MR450w 1.5T
Ge Medical Systems, LLC, Jun 9, 2016