Recall Z-0239-2019 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Philips Medical Systems Gmbh, DMC

Philips DigitalDiagnost software 4.1.x and 4.2.x, A Stationary X-ray System Models: 712027, 712028, 712029, 712031, 712032701032, 712033, 712224, 712225, 712226, 712227, 712228.

DateSep 19, 2018
StatusTerminated
ReasonStitching run on SkyPlate detector interrupted intermittently; Stitching with SkyPlate aborts after first image; System software CD contains wrong baseline
Root causeSoftware design
ActionPhilips issued letter dated 9/10/18 to advise users of the problem. Philips will install a system software version upgrade in affected systems, which will eliminate this issue. A Philips Service Engineer will contact account when the Field Action Kit is available to be implemented. Should you need to communicate with Philips with regard to this program, please reference Field Change Order 71200177.
Quantity829 units
DistributionWorldwide distribution - US Nationwide and countries of Canada Argentina Australia Austria Bangladesh Belgium Bolivia Brazil Chile China Colombia Czech Republic Denmark Egypt Finland France French Guiana Germany Guadeloupe Hong Kong Hungary India Indonesia Iran Israel Italy Japan Jordan Korea, Repu
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141736PHILIPS PROGRADE; PHILIPS ELEVA WORKSPOT FOR DIGITAL DIAGNOST
Philips Medical Systems Dmc GmbH, Jul 25, 2014