ClearanceBrief
Recall Z-0406-2019 ยท OSR Pacemaker/Icd/Crt Non-Implanted Components

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Medtronic CareLink 2090 Programmer

DateOct 2, 2018
StatusOpen, Classified
ReasonVulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
Root causeSoftware design
ActionThe firm began disseminating the Field Corrective Action notices by hand delivery on 10/02/2018 and followed with the posting of Medtronic Security Bulletin and the Urgent Medical Device Correction customer letter online on 10/11/2018. The notice explained vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download and beginning October 11th, 2018, Medtronic will be disabling the SDN for programmer updates and will rely solely on the USB update method.
Quantity30172 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.