Recall Z-2554-2019 ยท LKK Pump, Infusion, Implanted, Programmable
Smiths Medical ASD Inc.
Medfusion Model 4000 Syringe Infusion Pump, with PharmGuard Server Software (PGS). Sold under the following names: MEDFUSION 4000 Pump V1.0; MEDFUSION 4000 Pump V1.1, v1.1.1, v1.1.2; MEDFUSION 4000 Pump V1.1, V1.1.1, V1.1.2; MEDFUSION 4000 Pump V1.5, V1.5.1; MEDFUSION 4000 Pump V1.5, V1.5.1; MEDFUSION 4000 Pump V1.6, V1.6.1 Product Usage: The pumps are indicated for the following uses: 1. In
| Date | Oct 15, 2018 |
| Status | Terminated |
| Reason | Customers who utilize PharmGuard Server Software (PGS) with Medfusion 4000 pumps manufactured prior to November 10, 2017, may experience wireless network connectivity issues on certain pumps due to a potential Radio Module issue on some Medfusion 4000 pumps. |
| Root cause | Process control |
| Action | Smiths Medical notified customers on about 10/15/2018 via "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE". Specifically, Biomed individuals (owners) at each facility were notified because they are in contact with Smiths Medical service and repair team to conduct the correction. Instructions included the following actions: 1. To determine if your facility has any devices with a duplicate wireless MAC Address, a software script must be run on the PGS Database to collect MAC Address and serial number information. The PGS Database must be at version 2.0, 2.1, or 2.2 for this data collection proc |
| Quantity | 20,884 devices |
| Distribution | Worldwide distribution - Nationwide in the states of AL, AR, AZ, CA, CN, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV and country of Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K111386 | MEDFUSION MODEL 4000 SYRINGE INFUSION PUMP, PHARMGUARD TOOLBOX 2 MEDICATION SAFETY SOFTWARE AND SUPPORTED SYRINGES (PSS) Smiths Medical MD, Inc., Aug 29, 2011 |