Recall Z-0294-2019 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7, a stand-alone software treatment planning system.

DateSep 21, 2018
StatusTerminated
ReasonRobust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7.
Root causeSoftware design
ActionRay Search Labs notified userson 9/21/18 by email advising users of the action to be taken: After creating an adaptive plan from a plan with robust objectives, remove all robust objectives that are automatically added to the adaptive plan and replace them with other appropriate objectives before optimizing the adapted plan. o Do not use an optimization function template with robust functions for a dependent beam set.o Do not use scripting to add robust optimization functions related to beam set + background dose. Reply form to be completed/signed and returned. Regulatory information, please c
Quantity110 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180379RayStation 7
RaySearch Laboratories AB (PUBL), Apr 25, 2018