Recall Z-0513-2019 ยท DOP Mass Spectrometer, Clinical Use

PerkinElmer Health Sciences Canada, Inc.

PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC004872

DateSep 24, 2018
StatusTerminated
ReasonUnder certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.
Root causeSoftware design
ActionPerkin Elmer issued notification on 9/24/18 customers are requested to follow instructions contained within the Urgent Medical Device Correction letter and disseminate the information to appropriate functions within their organization. Inform relevant personnel and execute one of the following two options: Qption 1 (Users who do not require Autamatic Flushinq): Deactivate the automatic flushing option in your software: 1) Navigate to the Batch Acquisition Queue tab 2) Click the Queue Options button 3) Under the "System Flush and Standby" section, DESELECT the "Enable automatic flushing" optio
Quantity18 units
DistributionFL, MN, PA India, Italy, Korea, Spain