ClearanceBrief
Recall Z-0364-2019 ยท KPE Container, I.V.

Baxter Healthcare Corporation

EXACTAMIX Empty EVA Bags - 500 mL - Product Usage: A plastic container used to hold large-volume sterile solutions to be administered to a patient through an intravascular administration set.

DateNov 2, 2018 (initiated Sep 12, 2018)
StatusTerminated
ReasonPotential ability to leak once used for compounding.
Root causeProcess control
ActionAn Urgent Product Recall communication will be sent to affected customers via U.S.P.S., first class mail. Baxter is asking customers to: 1. Locate and remove all affected product from your facility. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. 3. Complete the enclosed Baxter Customer Reply Form and return it to Baxter by either fax or email. Notification Effectiveness: 100%. The non-responding customers will be contacted via telephone. Expansion on 02/19/2019, Baxter notified customers (and additional customers) of recall and additional lots included.
Quantity110,134 units
DistributionWorldwide distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. Countries of Canada, Australia, New Zeal
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K900585TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP
Baxa Corp., Sub. of Cook Group, Inc., Apr 18, 1990