Recall Z-0552-2019 ยท HAW Neurological Stereotaxic Instrument

Medtronic Navigation, Inc.

StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization StealthStation Cranial version 3.0, 3.1 CE 0344 Rx Only 2018-02 9735489 Revision 3 Medtronic Navigation, Inc. Biopsy Needle Kit (9733068) Note: This device is an accessory to your StealthStation System. . . The Biopsy Needle is a

DateDec 29, 2018 (initiated Sep 26, 2018)
StatusTerminated
ReasonAn incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.
Root causeDevice Design
ActionOn 09/26/2018, the firm sent an Urgent Medical Device Correction Notification to all US customers informing them of potential inaccuracy during biopsy procedures using the StealthStation S7 Depth Gauge feature on StealthStation S7 and i7 systems running Synergy Cranial and StealthStation Cranial software that could result in healthy tissue being biopsied. The notification letter stated, In navigated biopsy procedures, where the Navigate Projection feature is used and the projection is longer than the length of the plan, the graphical Biopsy Depth Gauge feature can display inaccurate informa
Quantity5487 impacted systems
DistributionUS: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY OUS: Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Ba
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150216Synergy Cranial
Medtronic Navigation, Inc., Jun 11, 2015
K153660StealthStation System with Cranial Software
Medtronic Navigation, Inc., Sep 14, 2016