ClearanceBrief
Recall Z-0435-2019 ยท FTM Mesh, Surgical

AROA Biosurgery, LTD.

OviTex 2S Reinforced BioScaffold 10x12cm, Part Number F10258-1012G

DateApr 6, 2018
StatusTerminated
ReasonDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
Root causeOther
ActionBeginning in April 2018, customers were visited by representatives and the affected units were replaced.
Quantity881 total
DistributionThe products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153633Endoform Reconstructive Template
Aroa Biosurgery Limited (Formerly Mesynthes Limited), Mar 15, 2016