Recall Z-0804-2019 ยท KTI Bronchoscope Accessory

Richard Wolf GmbH

Richard Wolf Bronchoscope Tubes, Model Texas. Device is reusable and should be sterilized prior to each use. Product Usage: Bronchoscopes are intended for investigation, diagnosis and therapy of pathological changes and removal of foreign bodies in the area of the tracheobronchial respiratory pathways, in conjunction with endoscopic accessories. The bronchoscopes can also be used endoscopically t

DateOct 17, 2018
StatusOpen, Classified
ReasonThere is a potential deficiency of the TEXAS Bronchoscope Tube product. The non-functionally relevant bonding points / cemented areas of the Bronchoscope Tubes may loosen.
Root causeProcess control
ActionOn about 09/24/2018, Richard Wolf GmbH sent Urgent Safety Information notices to international consignees. Customers in the USA were sent the "Urgent Safety Information" notice on about 10/17/2018. Customers were instructed to check current stock on hand for affected Bronchoscope Tubes, complete and return the attached response form, ensure all appropriate personnel at the facility are notified of the recall, ensure the recall information is observed in your respective institution until all internal measures have been completed, notify Richard Wolf of any adverse events, and return all affec
Quantity873 units worldwide (468 units within the US)
DistributionWorldwide distribution - US Nationwide in the states of PA, MD, NY, and countries of Bangladesh, Canada, China, France, Germany, India, Italy, Jordan, Netherlands, Poland, Russia, Spain, Thailand, Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.