Recall Z-0588-2019 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

GE Healthcare, LLC

GE LOGIQ E10 Ultrasound System, Model #530000E10. The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvagi

DateSep 14, 2018
StatusTerminated
ReasonThere is a potential for an undetected over temperature condition on the trans-esophageal (TEE) probe leading to a possible burn injury.
Root causeProcess change control
ActionThe recalling firm, GE Healthcare, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated 9/14/2018 via traceable means to each of their customer locations. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Do not use a TEE probe (GE model 6Tc-RS) from another device on the LOGIQ E10 system until your LOGIQ E10 is corrected. You may scan normally with any other probes. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. If you h
Quantity103 devices(64 US; 39 OUS)
DistributionWorldwide Distribution: US (nationwide) to states of: AZ, IL, MA, ME, MN, MT, PA, TX, and WA and countries of: Foreign distribution was made to Austria, Denmark, Finland, and Norway.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173555LOGIQ E10
GE Healthcare, Dec 20, 2017