Recall Z-0280-2019 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
Medtronic Navigation, Inc.-Littleton
O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027GERR SYS PRODUCT REFURB, Model Number Bl70000027GERR Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic
| Date | Aug 30, 2018 |
| Status | Terminated |
| Reason | Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems. |
| Root cause | Software design |
| Action | Urgent Medical Device Correction notification letters were sent to customers on 8/30/18. The letter identified the affected letter, problem and actions to be taken. All consignees will received the notification for 3.1.7 software maintenance release. Field service personnel will install the software onto the designed consignee systems. |
| Quantity | 596 total |
| Distribution | Worldwide - US Distribution in the states: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and Puerto Rico. The products were distributed to the following foreign countrie |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K092564 | O-ARM IMAGING SYSTEM Medtronic Navigation, Inc. (Littleton), Jun 17, 2010 |