Recall Z-0491-2019 ยท JTW System, Transport, Aerobic

Becton Dickinson & Company

BD Vacutainer Urine Collection Cups, Cat. No. 364941

DateAug 30, 2018
StatusTerminated
ReasonBD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
Root causeUnder Investigation by firm
ActionOn August 30, 2018, BD notified Sysmex via an emailed Urgent Medical Device Product Safety Alert letter. Sysmex was informed that BD had confirmed the product issue with BD Vacutainer Urine Collection Cups were used with Sysmex UF Series Analyzers. Sysmex was instructed to take the following actions: - Notify Sysmex customers of the potential for elevation of RBC counts - Complete the customer Resopnse form and email it to BD at Gail_Griffiths@bd.com to confirm receipt of the Safety Alert. Product will continue to ship so as to not impact critical patient care. BD recommends following Good
QuantityTotal: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS)
DistributionIllinois
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K790366URINE CULTURE KIT
Bd Becton Dickinson Vacutainer Systems Preanalytic, Apr 4, 1979