Recall Z-0400-2019 ยท NAE Dressing, Wound, Hydrogel Without Drug And/Or Biologic
Safe N Simple, LLC
Simpurity HYDROGEL ABSORBENT SHEET WOUND DRESSING, 4" X 5", 10 Pads, Catalogue number SNS58315
| Date | Aug 22, 2018 |
| Status | Terminated |
| Reason | The sterilization process used was unqualified. |
| Root cause | Process control |
| Action | Safe n Simple extended the recall to the only customer that ordered the product in the lot numbers (see table). The firm provided a written notice to the customer mentioned in the table with specific instructions along with a phone conversation. Safe n Simple instructed the customer to provide the inventory level of SNS58315 and SNS58345 specific to lot numbers 18060564 and 18060563 respectively and to return such product to Safe N Simple LLC at Safe N Simple LLC expense. The effectiveness of the recall was good since the customer immediately responded by shipping all their inventory back |
| Quantity | 104 cases |
| Distribution | CA |
| FDA record | Open on fda.gov |