Recall Z-0230-2020 ยท DQX Wire, Guide, Catheter
Stryker Neurovascular
Synchro2, 0.014in. x 200cm STANDARD PRE-SHAPED Guidewire with Hydrophilic Coating UPN Product No. M00326420, REF 2642, STERILE: EO - Product Usage: product is intended for general intravascular use, including neurovascular and peripheral vasculatures.
| Date | Oct 29, 2019 (initiated Jun 8, 2018) |
| Status | Terminated |
| Reason | The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error. |
| Root cause | Employee error |
| Action | On 6/08/18, "Urgent Medical Device Voluntary Customer Communication" were mailed via courier. The firm informed its customers that it has become aware that the product label (pouch and carton) for the affected products contain an incorrect value in millimeters (0.036 mm) for the Guidewire Outer Diameter dimension, but there will be no products being recalled. The Customer Notification Letter instructs customers to read this communication carefully and complete the following actions: 1. Immediately check your inventory. 2. Review the Communication and ensure full understanding of the contents. |
| Quantity | 6269 |
| Distribution | Worldwide distributions- U.S. Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, DE, FL, GA, GU, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY. Countries of Austria, Argentin |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K190843 | Synchro2 Support Guidewire Stryker, May 1, 2019 |