Recall Z-0102-2019 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
OriGen Biomedical, Inc.
Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO; and ECMO Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO
| Date | Aug 21, 2018 |
| Status | Terminated |
| Reason | The catheter failed the endotoxin testing. |
| Root cause | Release of Material/Component prior to receiving test results |
| Action | On 08/21/18, the firm emailed a recall notice to customers informing them of a nationwide recall of RDLC ECMO catheters. Customers were advised to return affected product to the firm. Customers with additional questions were advised to call 1-800-233-9014 or 1-512-474-7278. The firm also sent customers Technical Bulletin (TB) 1411 - Revision Feb 2018, Subject: Radiopacity of the Reinforced Dual Lumen Catheter (RDLC), Date: July 2015, and an Acknowledgment Receipt form to complete and return. On 09/12/18, and 09/20/18, the firm sent two additional recall notices, which included additional affec |
| Quantity | 106 |
| Distribution | US Nationwide in the states of NY, AR, MO, NC, GA, LA, CO, AL, MA, CA, MI, TX, OH, MD, FL, TN, SC, OR, AZ, UT, RI, SD, PA, NM, CT. Foreign (OUS): United Arab Emirates, Kingdom of Saudi Arabia, New Zealand, Thailand |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K113869 | ORIGEN REINFORCED DUAL LUMEN CATHETER Origen Biomedical, Inc., Jul 26, 2012 |