Recall Z-0462-2019 ยท DQK Computer, Diagnostic, Programmable

Invivo Corporation

Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

DateSep 24, 2018
StatusCompleted
ReasonThe Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard
Root causeSoftware Design Change
ActionCustomers were notified on about 09/20/2018 via an "URGENT - Medical Device Correction" letter. An initial letter was sent to distributors with instructions to provide the Field Safety Notice to all customers the devices were further distributed to. The "URGENT - Medical Device Correction" letter provided guidance on how to identify the Service Number and serial number on the Flex Cardio device, which is on the bottom right corner of the back of the device. Actions to be taken by the customer/user until Philips contacts the customer to arrange for a software update to address this issue,
Quantity76 units
DistributionDistributed to the Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101571XPER INFORMATION MANAGEMENT AND FLEX CARDIO PHYSIOMONITORING SYSTEMS
Witt Biomedical Corp, Oct 26, 2010