Recall Z-0065-2019 ยท IYE Accelerator, Linear, Medical

Vision RT Ltd

Integrated Gate Controller PCB (component of AlignRT Plus) Truebeam variant only. The AlignRT is a video-based three-dimensional (3D) surface imaging system which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. Part number: V000172

DateAug 24, 2018
StatusTerminated
ReasonRTInterface error in the event of an I/O error between the IGC PCB and the Framegrabber PCB
Root causeSoftware design
ActionVision RT notified customers with Customer Information Bulletin dated August 24, 2018 to make them aware of the issue, how to spot it (in the unlikely event that it occurs) and how to deal with it. The customer contact also indicates that the issue is to be fixed with a firmware update and that this can be delivered remotely or installed on site by a service engineer,or transmit the firmware update to those customers whose systems can be accessed .
Quantity690 systens
DistributionNationwide Foreign: Australia, Austria, Azerbaijan, Belgium, Brunei, Bulgaria, Canada, China, Croatia, Denmark, Finland, Georgia, Germany, India, Israel, Italy, Japan, Korea, Lithuania, Malaysia, Morocco, Norway, Poland, Portugal, Puerto Rico, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123371ALIGNRT PLUS
Vision RT Limited, Jan 8, 2013