Recall Z-0291-2019 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Stryker GmbH

Hoffman LRF Hexapod Strut, Extra Short 183mm-295mm, Model Number 4935-0-040. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of

DateAug 28, 2018
StatusTerminated
ReasonThe press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
Root causeUnder Investigation by firm
ActionUrgent Product Recall notification letters dated 8/28/18 were sent to customers.
Quantity867
DistributionUS Nationwide in the states of CO, FL, MD, MI, NY, OK, and PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163656Hoffmann LRF System
Stryker GmbH, Apr 14, 2017