Recall Z-0584-2019 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

PROCARE B40 Patient Monitor, V1; and B40 Patient Monitor, V2, V2.1, and V3. The PROCARE Monitor B40 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.

DateSep 11, 2018
StatusOpen, Classified
ReasonPatient monitors may restart due to network overload caused by network configuration.
Root causeSoftware design
ActionThe recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 9/11/2018 via trackable mail on 9/11/2018 to its customers. The letters described the product, problem and actions to be taken. The letters informed the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Bx0 and/or B1x5 patient monitor on the network does not resume normal functionality in approximately 90 seconds. The customers were instructed to follow the safety instructions. GE Healthcare will provide revised network
Quantity33,065 patient monitors
DistributionWorldwide Distribution: US (nationwide) including Puerto Rico; and countries of: Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Brunei, Bulgaria, Burma (Myanmar), Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republ
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120598PROCARE MONITOR B40
Ge Medical Systems China Co., Ltd., Jul 19, 2012
K130584MONITOR B40
Ge Medical Systems China Co., Ltd., Jun 26, 2013
K133576MONITOR B40
Ge Medical Systems China Co., Ltd., Feb 21, 2014
K151063Monitor B40
Ge Medical Systems China Co., Ltd., Dec 15, 2015