Recall Z-0096-2019 ยท DQK Computer, Diagnostic, Programmable

GE Healthcare Finland Oy

GE Healthcare CARESCAPE Monitor B850 Product Usage: Intended uses of CARESCAPE B850 V2 in 510(k) countries [UR_CRSCP-1046]: The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time.

DateSep 4, 2018
StatusOpen, Classified
ReasonWhen multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
Root causeSoftware design
ActionThe firm issued the Medical Device Correction Notice on 09/04/2018 by letter. The letter explained the issue and provided the following safety instructions: If a prolonged network overload occurs and CARESCAPE Monitor Bx50 on the network does not resume normal functionality in approximately 90 seconds, take the following actions: 1. Temporarily switch the monitor from central monitoring to local-only (bedside) monitoring: - Disconnect the LAN cables from IX/MC ports on backside of the monitor. The monitor will now be in local-only mode. - Keep the patient under close observation and con
Quantity47,937 units
DistributionWorldwide Distribution-US Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132533CARESCAPE MONITOR B450
GE Healthcare Finland Oy, Dec 4, 2013