Recall Z-0522-2019 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

Solar 8000i Patient Monitoring System is used to monitor physiologic parameter data on adult, pediatric and neonatal patients.

DateAug 31, 2018
StatusOpen, Classified
ReasonThe patient monitors may simultaneously restart as designed if all are connected to the same network and a network overload occurs for a prolonged time.
Root causeSoftware design
ActionThe recalling firm issued letters dated 8/31/2018 via FedEx on 8/31/2018 informing the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Dash and/or Solar patient monitor on the network does not resume normal functionality in approximately 90 seconds.
QuantityApproximately 73,673 Solar 8000i and 8000M monitoring systems
DistributionDistribution was made nationwide, including Puerto Rico, Guam, and Samoa. There was also government and military distribution. Foreign distribution was made to Canada, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosn