Recall Z-0296-2019 ยท PAM Gram-Positive Bacteria And Their Resistance Markers

BioFire Diagnostics, LLC

FilmArray Blood Culture Identification (BCID), IVD, Rx Only, BioFire Diagnostics, LLC, Panel Catalog Numbers/ UDI: RFIT-ASY-0126 (30 test kit)/ 00815381020086, RFIT-ASY-0127 (6 test kit)/ 00815381020093, when used with the following blood culture bottles: Bottle Description/Part Number: BACT/ALERT FA Plus/ 410851, BACT/ALERT FN Plus/ 410852, BACT/ALERT PF Plus/ 410853,

DateOct 30, 2018 (initiated Sep 5, 2018)
StatusTerminated
ReasonIncreased risk of false positive results for Proteus when the product is used with a specific blood culture bottles.
Root causeMaterial/Component Contamination
ActionOn 09/05/2018 the firm sent an Urgent Customer Advisory Notice to all customers via Federal Express (to US customers) and USPS (to Laboratory Directors for all organizations that have ordered kits) informing them that the firm has identified an increased risk of false positive Proteus results when the BioFire BCID Panel is used with BACT/ALERT Blood Culture Bottles: BACT/ALERT Blood Culture Bottle Catalog No. Description 410851 BACT/ALERT FA Plus 410852 BACT/ALERT FN Plus 410853 BACT/
Quantity18038
DistributionWorldwide distribution. US nationwide, Kuwait, Lebanon, Malaysia, Myanmar, Nicaragua, Oman, Panama, Paraguay, Peru, Philippines, Qatar, Romania, Saudi Arabia, Slovenia, Uruguay, United Arab Emirates, Vietnam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130914FILMARRAY BLOOD CULTURE IDENTIFICATION (BCID) PANEL
Biofire Diagnostics, Inc., Jun 21, 2013
K143171FilmArray Blood Culture Identification (BCID) Panel for use with the FilmArray 2.0
Biofire Diagnostics, LLC, Jan 30, 2015
K160457FilmArray Blood Culture Identification (BCID) panel for use with FilmArray Torch
Biofire Diagnostics, LLC, Mar 15, 2016
K181493FilmArray Blood Culture Identification (BCID) Panel
Biofire Diagnostics, LLC, Jul 5, 2018