Recall Z-0208-2019 ยท LXH Orthopedic Manual Surgical Instrument

DePuy Mitek, Inc., a Johnson & Johnson Co.

Mitek Screw and Washer Depth Gauge, Product Code 219024

DateAug 30, 2018
StatusTerminated
ReasonThe device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.
Root causeUnder Investigation by firm
ActionUrgent Medical Device Recall Notification letters were sent to customers on 8/30/18. Updated Urgent Medical Device Recall notification letters were sent to customers on 9/18/18
Quantity401
DistributionUS nationwide in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MI, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WA, WI. The products were distributed to the following foreign countries: Australia, Canada, Chile, China, Chechnya, Denmark, Italy, Japan
FDA recordOpen on fda.gov