Recall Z-0118-2019 ยท LLZ System, Image Processing, Radiological

AS Software, Inc.

AS-OBGYN Information System version 7.824.x

DateSep 20, 2018
StatusTerminated
ReasonAny change made to current ultrasound machine software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, must be tested prior to the use in production environment to ensure that ultrasound measurements are crossing over correctly to the AS system
Root causeDevice Design
ActionAS Software issued letter by email on 9/20/18 to notify customers that any change made to their ultrasound machine such as software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, are tested prior to the use in the production environment. Medical Device Correction Return Response form to be signed and returned. The AS Software Support Department can be reached Monday through Friday, 9:00 AM to 6:00 PM EST via phone: 201-541-1900 option 2 or e-mail: info@as-software.com
Quantity52 systems
DistributionUS Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051639AS-OBGYN INFORMATION SYSTEM; AS-EMR SYSTEM
As Software, Inc., Aug 2, 2005