Recall Z-0070-2019 ยท GAG Stapler, Surgical

COVIDIEN LLC

EEA Circular Stapler with Tri-Staple Technology, Part No. TRIEEA28MT TRIEEA28XT TRIEEA31MT TRIEEA31XT Product Usage: The EEA Circular Stapler with Tri-Staple Technology places a circular, triple staggered row of titanium staples and resects the excess tissue, creating a circular anastomosis.

DateAug 17, 2018
StatusTerminated
ReasonThe firm identified the potential for a device to have an incorrect tissue gap. Use of a device with an incorrect tissue gap may result in incomplete staple formation and/or the inability to remove the device from tissue following application potentially leading to bleeding, anastomotic leak or tissue trauma. There have been no reports of serious injury related to this issue.
Root causeProcess design
ActionAll U.S. consignees were notified via Federal Express or certified mail on August 17, 2018. The letter informs customers of the potential for use of a device to have an incorrect tissue gap. Required Actions: 1. Please quarantine and discontinue use of the affected item code and lots listed above. 2. Please return affected product. All unused products from the affected item codes and lots must be returned. If you purchased directly from Medtronic, return all inventory of the affected lots to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department.
Quantity1,902
DistributionWorldwide Distribution - US Nationwide in the states of CA, FL, IL, NC, NY, UT. and the countries of Austria, Belgium, Denmark, Finland, France, Germany, Iceland, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Reunion, Spain, Sweden, Switzerland, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172361EEA Circular Stapler with Tri-Staple Technology
Covidien, Feb 16, 2018