ClearanceBrief
Recall Z-0143-2019 ยท HSB Rod, Fixation, Intramedullary And Accessories

Zimmer Biomet, Inc.

Affixus Hip Fracture Nail Right 125 13 mm x 400 mm, Item Number 814313400 Product Usage: Intended for the fixation of fractures

DateAug 8, 2018
StatusTerminated
ReasonThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Root causeEmployee error
ActionUrgent Medical Device Recall notification letters dated 8/8/18 were sent to customers. The letters identified the affected products, problem and actions to be taken. Customers were instructed to: 1. Complete the Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 2. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call ce
Quantity10
DistributionWorldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WY. The products were distributed to the following foreign countries: Chile,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100238HFN 125 DEGREE 9-13 MM, 180MM, HFN 130 DEGREE 9-13 MM, HFN RH 125 DEGREE 9-13MM
DePuy Orthopaedics, Inc., May 11, 2010