Recall Z-0199-2019 ยท LNH System, Nuclear Magnetic Resonance Imaging

Canon Medical System, USA, INC.

Nuclear Magnetic Resonance Imaging Systems that are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body.

DateMay 30, 2018
StatusOpen, Classified
ReasonWhen using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value. The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.
Root causeDevice Design
ActionOn 05/20/2018 the firm sent an Urgent Medical Device Correction letter to customers via United States Postal Service return receipt mail informing them that when a user used the respiratory gating with Steady Mode ON in Spin Echo sequence scan, the SAR value displayed on console was lower than the actual SAR value because there was a wrong estimating the RF pulses in the software. The SAR may exceed the upper limit, when a user scans a patient near the upper limit of the first level controlled operating mode in case of using the respiratory gating with Steady Mode ON in Spin Echo sequence sca
Quantity252
DistributionDomestic: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY. International: Algeria, Australia, Bangladesh, Brazil, Canada, China, Colombia, Congo, Dominican Republic, Egypt, Guinea,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120638VANTAGE TITAN
Toshibamedical Systems Corporation, Jun 1, 2012
K170412Vantage Titan 1.5T, MRT-1510, M-Power GX
Toshibamedical Systems Corporation, Jun 7, 2017
K171597Vantage Elan 1.5T, V4.0
Toshibamedical Systems Corporation, Jul 21, 2017