Recall Z-0004-2019 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems

TrueBeam Radiotherapy Delivery System version 2.0 and 2.5 EDGE Radiotherapy Delivery System version 2.0 and 2.5 Product Usage: The TrueBeam delivery systems are intended to provide radiotherapy and stereotactic radiosurgery for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

DateAug 20, 2018
StatusTerminated
ReasonReports have been received of an anomaly that can result in a treatment without intended gating (respiratory tracking/monitoring). This issue occurs when a patient planned with gating is treated on more than one system.
Root causeSoftware design
ActionOn 08/20/2018 the firm sent an URGENT MEDICAL DEVICE CORRECTION FIELD SAFETY NOTICE to customer via courier service (US and Canada via FedEx on 8/23/2018, and Rest of World via third party courier on 8/24/2018) informing them the firm had received reports of an anomaly that can result in a treatment without intended gating, and that this issue occurs when a patient planned with gating is treated on more than one system. Customers were advised, for treatment plans with intended gating, the firm recommends users monitor and pay attention to the system status message. When a patient with a gated
Quantity387
DistributionWorldwide Distribution - US nationwide in the states of AL, AR, AZ, CA, CO, FL, Guan, IL IN, KS, KY, LA, MA, MD, MI, MO, NH, NJ, NY, OH, OK, PA, Puerto Rico, TN, TX, VA, WI and the countries of : Algeria, Australia, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, China, Croatia, Czech Republic, Denma
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140528TRUE BEAM, TRUE BEAM STX, EDGE
Varian Medical Systems, Inc., Sep 5, 2014
K143224TrueBeam-TrueBeam STx-Edge
Varian Medical Systems, Inc., Dec 19, 2014