Recall Z-0359-2019 ยท JQP Calculator/Data Processing Module, For Clinical Use
Beckman Coulter Inc.
Sorting-Drive Professional Part, Sorting-Drive Professional Kit, (Sorting Drive Software v4.2.1), Part No. A94144, Beckman Coulter, Inc. The Sorting-Drive Professional Kit (PN A94144) includes the Sorting-Drive 4.2.1 User Kit (PN B95347) which includes the Sorting-Drive 4.2.1 Software DVD (PN B95346). This software version includes an option for manual barcode printing for sample tubes. Sortin
| Date | May 24, 2018 |
| Status | Terminated |
| Reason | Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes.Sorting Drive software version 4.2 was replaced by 4.2.1 (both versions contain the same option for manual barcode printing). When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports that this does not affect automated label creation or reprinti |
| Root cause | Software change control |
| Action | On 6/4/2018 Beckman Coulter (BEC) reported that the issue will be corrected via a field modification performed by Field Service. BEC reported that they do not intend to send a customer letter for this issue.US customers can contact the Technical Support Troubleshooting Hotline at (800) 854-3633. |
| Quantity | 58 |
| Distribution | Worldwide Distribution: US (nationwide) and countries of: Algeria, Argentina, Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland, Israel, Italy ,Republic of Korea, Lebanon, Lithuania ,Malaysia, Netherlands, Poland, Portugal , Qatar |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.