ClearanceBrief
Recall Z-3255-2018 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific Corporation

ESSENTIO Pacemaker

DateSep 10, 2018
StatusTerminated
ReasonBoston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.
Root causeComponent design/selection
ActionOn September 2018 Boston Scientific sent letters to their customers with the following information: Dear Doctor, Boston Scientific has identified a subset of approximately 2,900 active ACCOLADE", PROPONENT", and ESSENTIO" pacemakers and VISIONIST" and VALITUDE" cardiac resynchronization therapy pacemakers (CRT-Ps) with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion. There have been no reports of injury related to this behavior. You are receiving this letter because you or your center has implanted or follow one or more patients with
DistributionAK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, W A, WI, WV, and WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.