Recall Z-0108-2019 ยท HJO Biomicroscope, Slit-Lamp, Ac-Powered

Carl Zeiss Meditec AG

Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 1185393 2017-09-01

DateJun 25, 2018
StatusTerminated
ReasonCalibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
Root causeEmployee error
ActionOn 08/06/2018 the firm sent a Field Safety Corrective Action letter to customers via FedEx Priority Overnight service informing them that based on routine post market surveillance activities, we have information that this specific IOLMaster 700 device is incorrectly calibrated. The calibration value for axial length measurement was not entered correctly in the device settings during the manufacturing process. The transmission discrepancy leads to an error during the axial length measurement of approx. 0.25mm. A deviation in the axial length of this magnitude results in an error of Single unit.
Quantity1
DistributionUS: CA (Mountain View and San Jose) A single system. No distributed OUS.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170171IOLMaster 700
Carl Zeiss Meditec, AG, Apr 6, 2017