Recall Z-3175-2018 ยท EPT Microscope, Surgical

Leica Microsystems, Inc.

Leica M525 F20 Surgical Microscope System

DateSep 14, 2018 (initiated Sep 4, 2018)
StatusTerminated
ReasonThe firm received complaints that the instrument tipped over when the swing arm was fully extended.
Root causeDevice Design
ActionThe firm disseminated their Field Corrective Action Notices by mail beginning on 09/04/2018 to the US and Canadian consignees. The letter explained the problem and informed the consignee that an update package with enhanced instructions for use and additional marking will be provided within the next 3 months. Additional consignees found were notified by email on 10/01/2018.
Quantity160 units
DistributionUS and Canada