Recall Z-0062-2019 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Dannoritzer Medizintechnik GmbH & Co. KG

Laparoscopy Instruments Instructions For Use (IFU) for the following: 1)Monopolar electrode, 2) Insert, handle and/or shaft component of a modular forceps or scissors for electrosurgical coagulation, 3) One-piece forceps or scissors for electrosurgical coagulation

DateAug 6, 2018
StatusTerminated
ReasonThe Manual cleaning process and the usability of the reprocessing instruction by the user had not been validated. The IFU was modified to remove the manual cleaning method and revised the instructions for use for clarity and readability.
Root causeLabeling False and Misleading
ActionIn July 2017, Dannoritzer Medizintechnik GmbH & Co. KG issued emails to customers notifying them that an updated Instructions For Use (IFU) manual would be distributed by July 27, 2018. On August 8, 2018, Dannoritzer Medizintechnik GmbH & Co. KG issued Urgent Medical Device Field Correction notices dated August 6, 2018 to customers along with an updated IFU. Customers were encouraged to immediately remove from their place of use and destroy the existing original DANmed Laparoscopy Instruments IFU as well as any copies and replace them with the revised version (V.2018-08-03) included with the
DistributionUS Nationwide in the States of OH, MI, MA, CA, OK, TN, GA, FL, IL, NC, TX, MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052759DAN MONOPOLAR LAP ACCESSORIES
Dannoritzer Medizintechnik GmbH & Co. KG, Jun 14, 2006