Recall Z-0067-2019 ยท DSQ Ventricular (Assist) Bypass

Heartware, Inc.

Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation

DateAug 28, 2018
StatusTerminated
ReasonCertain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.
Root causeDevice Design
ActionMedtronic began notifying Clinicians in the US on 28-Aug-2018 mainly via 2-day courier, of the identified issues and provided patient management recommendations. Outside the US, Medtronic began notifying Clinicians on 30-Aug-2018 via locally approved methods.
Quantity17,275 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.